Artifact Correction for Echo-Planar Imaging at Low-Field and Ultra-Low-Field MRI
About
Purpose: Echo-planar imaging (EPI) in low-field (LF) and ultra-low-field MRI (ULF) suffers from severe Nyquist ghost artifacts due to odd-even k-space misalignment. This study develops a reference-free artifact correction pipeline that reduces reliance on conventional reference scans while achieving improved ghost suppression. Methods: Starting from the traditional reference-scan-based ghost artifact correction method, we first introduce a peak-alignment-based ghost artifact correction method to correct odd-even line displacement without reference data. To further reduce residual artifacts, an interpolation-and-resampling strategy is applied. The combined method was evaluated using EPI and diffusion-weighted EPI data in LF and ULF. Results: The proposed pipeline effectively mitigated Nyquist ghosts, improved structural continuity, and enhanced signal uniformity. Peak-alignment-based ghost artifact correction method alone provided comparable artifact suppression to reference-scan-based ghost artifact correction method, while interpolation and resampling further suppressed residual artifacts, enabling reliable visualization of brain structures under ULF conditions. Conclusion: A practical, reference-free correction pipeline is presented for LF and ULF EPI, combining peak-alignment-based ghost artifact correction method and interpolation-resampling to achieve efficient ghost suppression and expand the clinical applicability of low-field MRI systems, providing both theoretical guidance and practical experience for ULF EPI-based DWI imaging.
Related benchmarks
| Task | Dataset | Result | Rank | |
|---|---|---|---|---|
| DW-EPI ghost artifact correction | Phantom at b = 0 s/mm² | GSR0.046 | 5 | |
| DW-EPI ghost artifact correction | Phantom at b = 400 s/mm² | GSR0.1304 | 5 | |
| DW-EPI ghost artifact correction | Volunteer at b = 0 s/mm² | GSR3.88 | 5 | |
| DW-EPI ghost artifact correction | Volunteer at b = 400 s/mm² | GSR0.0618 | 5 | |
| EPI ghost artifact correction | Phantom at 0.5 T MRI system | GSR0.0172 | 5 | |
| EPI ghost artifact correction | Phantom 0.068 T MRI system | GSR0.0454 | 5 |